PAT-Power Analyser Transfer
Download
The PAT Power Analyser Transfer software can be used to control the single-phase wattmeters and energy analysers, as well...
I want to use my F607 clamp on a 10x80 busbar. Is this possible?
FAQ
Yes, this is possible with the F607 clamp.
You will find drawings showing the dimensions accepted by the F40x and F60x multimeter/wattmeter...
Android application for F407-F607
Download
F407 - F607 Android application We have an Android application available for the F407 and F607 power and harmonics...
How can I set the Inrush trigger level with the F407 and F607 clamps?
FAQ
At startup, press MAX/MIN + A to modify the value of the current detection threshold for INRUSH measurement. The default values are 10%, i.e. 100 A...
How can I adjust the measurement acquisition interval on the F407 and F607...
FAQ
At startup, press 2nd + W. This will modify the recording interval in the EEPROM. The default value is 60 and the possible values range from 1 second...
How can I erase the memory of the F407 and F607 clamps?
FAQ
To delete recordings from the EEPROM (Electrically Erasable Programmable Read Only Memory), press 2nd + Ohm when you power up the clamps. No...
How can I perform a Factory Reset on the F407 & F607 clamps?
FAQ
To return to the default configuration, enter the following Key-Switch combination: 2nd + A
What is the Bluetooth communication code of the F607 and F407 to be entered...
FAQ
For the first connection, you must enter the code 0000 in the Bluetooth manager window. Communication is then established automatically.